Instruction Manual Mi 180 Bench Meter
3 8
3 8
3 8
3 8
3 8
PC INTERFACE
Data transmission from the instrument to the PC can be done with the Mi5200 Windows
®
compatible software, when using the RS232 or USB serial interfac
e
. Mi Mi5200 also offers
graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for further analysis.
To connect the instrument to a PC through the RS232 port, use the M
ilwaukee
MA9350
cable
connector.
To connect the instrument to a PC through the USB port, use a standard USB cable.
Make
sure that your instrument is switched off and plug one connector of the cable to the
instrument RS232/USB connector and the other to the serial port of your PC.
Notes
Notes
Notes
Notes
Notes:
•Other cables than MA
MA
MA
MA
MA 9350
9350
9350
9350
9350 may use a different configuration. In this case, communication
between instrument and PC may not be possible.
•Keep only one cable connected (RS232 or USB) during PC communication to avoid
possible errors.
•If you are not using Mi5200 software, please see the following
instructions.
SENDING COMMANDS FROM PC
SENDING COMMANDS FROM PC
SENDING COMMANDS FROM PC
SENDING COMMANDS FROM PC
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal program. Use MA 9350
MA 9350
MA 9350
MA 9350
MA 9350
cable to connect the instrument to a PC, start the terminal program and set the communica-
tion options as follows: 8, N, 1, no flow control.
COMMAND TYPES
COMMAND TYPES
COMMAND TYPES
COMMAND TYPES
COMMAND TYPES
To send a command to the instrument follow the next scheme:
<*> <command
>
<CR>
where: <*> is the command prefix.
<command> is the command code.
Note
Note
Note
Note
Note: Either small or capital letters can be used.
UNIT CHANGE COMMAND
UNIT CHANGE COMMAND
UNIT CHANGE COMMAND
UNIT CHANGE COMMAND
UNIT CHANGE COMMAND
CHU xx
CHU xx
CHU xx
CHU xx
CHU xx Change the instrument unit according with the parameter value (xx):
• xx=00 pH range/0.001 resolution
• xx=01 pH range/0.01 resolution
• xx=03 mV range
• xx=04 Rel mV range