•If the reading is out of range, the full-scale value (160.0 g/L for
MTC
/
ATC
mode or
400.0 g/L for uncompensated TDS) will be displayed blinking.
•If the SHIFT&RANGE/FIXED keys are pressed to freeze the LCD range and the reading
goes out of range, the full-scale value of the frozen range will be displayed blinking.
NaCl
NaCl
NaCl
NaCl
NaCl MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
• Press the MODE/SETUP key while in EC range to enter NaCl measurement mode.
• The instrument will display the NaCl reading on the primary LCD and the temperature on the
secondary LCD, along with the reference temperature.
AUTORANGING
The EC and TDS scales are autoranging. The meter automatically sets the scale with the
highest possible resolution.
By pressing the SHIFT&RANGE/FIXED keys, the autoranging feature is disabled and the
current range is frozen on the LCD. The “Auto” “OFF” (autoranging disabled) message
will be displayed on the LCD for a few seconds. To restore the autoranging option,
press the SHIFT&RANGE/FIXED keys again. The “Auto” “On” (autoranging enabled)
message will be displayed on the LCD for a few seconds.
Note
Note
Note
Note
Note: Autoranging is automatically restored if the range is changed, if the setup or
calibration modes are entered and if the meter is turned off and back on again.