Instruction Manual Mi 180 Bench Meter
3 4
3 4
3 4
3 4
3 4
If there is another item to be set (e.g. minutes), press the RANGE/FIXED key, and that item will start
blinking.
Press the arrow keys to change the blinking value.
Press the GLP/ACCEPT key to accept the value or the CAL key to escape.
Press the arrow keys to select the next/previous parameter.
Press the SHIFT
&
MODE/SETUP key to exit SETUP menu at any time.
The following table lists the SETUP parameters, their valid values range and the factory settings
(default):
Note
Note
Note
Note
Note: If “dISP” option is ON, the tags corresponding to the calibrated buffers are displayed on
the LCD while in pH measurement mode.
Item
Description
Valid values
Default
OFF dAY
Alarm Time Out
OFF or 1 to 14 days
OFF
dISP
Display Cal Buffers
ON/OFF
ON
CELL
Cell Constant
0.500 to 1.700
1.000
tdS
TDS Factor
0.40 to 0.80
0.50
tc
Temperature Coefficient
0.00 to 6.00%/°C
1.90
rEF
Reference Temperature
20.0 or 25.0 °C
25.0
TIME
Time (hh:mm)
00:00 to 23:59
00:00
DATE
Date (MM.DD.YYYY)
01.01.2000-12.31.2099 01.01.2005
bEEP
Beep Status
ON/OFF
OFF
bAud
Baud Rate
600; 1200; 2400; 4800;9600
2400
In Id
Instrument ID
0000 to 9999
0000
tEMP
Temperature Unit
°C or °F
°C