Instruction Manual Mi 180 Bench Meter
2 4
2 4
2 4
2 4
2 4
LAST
LAST
LAST
LAST
LAST NaCl
NaCl
NaCl
NaCl
NaCl CALIBRATION DATA
CALIBRATION DATA
CALIBRATION DATA
CALIBRATION DATA
CALIBRATION DATA
Last NaCl calibration data is stored automatically after a successful calibration. To view the NaCl
calibration data, press the GLP/ACCEPT key when the instrument is in NaCl measurement mode.
The instrument will display the time (hh:mm:ss) of the last calibration as in pH GLP mode.
Press the UP and DOWN arrow keys to view the next logged calibration parameters
(pressing the UP key):
• The date (MM.DD.YYYY) as in pH GLP mode.
• The cell constant.
• The salinity coefficient.
• The temperature coefficient on the primary LCD, along with the temperature compensation
mode as in EC GLP mode.
• The reference temperature on the primary LCD as in pH GLP mode.
• The instrument Identification Code as in pH GLP mode.
Notes
Notes
Notes
Notes
Notes:
• If a one-point pH calibration is performed after a two-point pH calibration, the
instrument will keep the old slope.
• Press the GLP/ACCEPT key at any moment and the instrument will return to measure-
ment mode.