Sterylis VS Manual
| 35
The device can communicate with the user through the following indications:
Indication
Meaning
Green filtration mode light is on
The device works in filtration mode
Green silent mode light is on
The device works in silent disinfection mode
Green standard mode light is on
The device works in standard disinfection mode
Countdown of the time on the display in the
OZONING TIME section
The device, after a timeout, will go into sterilization mode
or
The device will stop working in the sterilization mode after
deduction of the indicated time (AUTO sterilization mode)
or the ozone generators will be switched off after
deduction of the indicated time and the device will go to
ozone destruction (MANUAL sterilization mode)
Flashing green light in filtration mode
PRE-TEST function for filtration mode in progress, no
active errors
Flashing green light in silent mode
PRE-TEST function for silent disinfection mode in
progress, no active errors
Flashing green light in standard mode
PRE-TEST function is in progress for standard disinfection
mode, no active errors
Flashing green light in filtration mode + flashing
text tESt
PRE-TEST function for filtration mode in progress, active
errors
Green silent mode light bl tESt text
blinking
PRE-TEST function for silent disinfection mode is running,
active errors
Standard mode green light bl tESt text
blinking
PRE-TEST function is in progress for standard disinfection
mode, active errors
Yellow pictograms O
3
and ! flashing
PRE-TEST function for sterilization mode in progress, no
active errors
Yellow pictograms O
3
and ! fl tESt
flashing
PRE-TEST function for the sterilization mode in progress,
active errors
Safe ozone concentration light is on
Ozone concentration in the room is safe for the user (<0.1
ppm)
Safe ozone concentration light is off (despite the
lack of ozone in the room)
Ozone sensor error detected
Safe ozone concentration flashing light
Ozone sensor heating process is in progress
TESt blinking
AUTO-TEST function in progress
O3_t blinking
OZONE-SENSOR-TEST function in progress
XXXX or HHHH text display and short beep
Incorrectly entered access code
Display the L0CH message
The device is in lock mode
The display in the CONCENTRATION section
shows the current ozone concentration, the
work of the beacon, the rest of the operator
panel is dimmed
The device is in sleep mode, an ozone concentration
exceeding the safe level for the user has been detected
All displays blink and a short beep
Storing the value of a selected configuration parameter or
Resetting the value of the selected counter or
Summoning the selected service function
Flasher beacon operation
The measured ozone concentration exceeding 0.1 ppm or
Active sterilization process or
Integration tests F004 and F005
Buzzer operation
Active sterilization process
Backlit panel operation
Operation of UV-C lamps (silent disinfection mode,
standard disinfection or ozone destruction active)