background image

 

 

 

Sterylis VS Manual                                                                                                                                                                                       

| 32

 

 

The active carbon filter that is pulled out of the destination causes a violation of the limiting switch. If a limiting switch is 
violated during the sterilisation process, the process will be stopped and completed and the device will go into standby 
mode. In case of violation of the limiting switch in filtration or disinfection mode, the process will be stopped at the current 
moment and the unit will go into pause mode. When the limiting switches of the filter cartridge are no longer violated 
(inserting the filter cartridge back), the unit will resume the filtration/disinfection process. Violation of the end point also 
results in a corresponding error on the display in the O

3

 STERILIZATION section. If the end point violation stops, the error is 

automatically reset. 
 
The safety limiting switches provide protection in case of an attempt to change the air filters during operation.  

7.3 Replacing the UV-C lamps 

Regular replacement of the UV-C radiators is essential for proper operation. Exceeding the service life of UV light bulbs 
may  result  in  their  burning  out.  It  should  also  be  remembered  that  with  time  the  maximum  dose  of  lamp  radiation 
decreases.  
 
The lifetime of UV bulbs is approx. 9000 hours. This is at the same time the time after which a message indicating that 
this time has been exceeded will be displayed.  
 

 
Note! The replacement of UV-C fluorescent tubes in STERYLIS sterilisers may only be carried out by the 
manufacturer's service or another authorised service centre! 

 
 

7.4 Replacing the ozone generators 

Regular replacement of the UV-C lamps is essential for proper operation. The lifetime of the generators is approx. 9000 
hours. 

 
 
Note!  The  replacement  of  ozone  generators  in  STERYLIS  sterilisers  may  only  be  carried  out  by  the 
manufacturer's service or another authorised service centre. 

 
 

7.5 Ozone sensor service 

The  ozone  sensor  is  one  of  the  most  important  elements  of  the  device,  which  determines  its  proper  functioning  and 
safety. Regular recalibration and replacement after the expiry date of the ozone sensor is essential for proper and safe 
operation.  Re-calibration  should  take  place  every  1,000  hours  of  operation  of  the  device  in  sterilization  mode,  while 
replacement is related to the expiry date after 5 years from the purchase of the device. 
Messages for recalibration and expiry date of the ozone sensor are displayed by means of codes  F004 and F005 (See: 
Chapter 9) and the warning beacon. When the counters are reset after recalibration or sensor replacement (according to 
the instructions in Chapter 6.5), the signaling by means of messages and the warning beacon is switched off. 

 
 
Note! The replacement and re-calibration of the ozone sensor in STERYLIS sterilizers may only be carried 
out by the manufacturer's service or another authorised service centre

 
 

 
 
 

Содержание STERYLIS VS-100/450

Страница 1: ...MODELS VS 100 450 VS 160 600 VS 300 900 VS 400 1200 VS 600 1500 VS 900 1800 MANUAL USER S PROFESSIONAL ROOM STERILISERS...

Страница 2: ...with the ozone concentration limits 23 6 3 4 5 Emergency situations during the sterilization process 24 6 4 Sleep mode 24 6 5 Front panel locking mode 24 6 6 Changing the configuration parameters 25...

Страница 3: ...inside the device is high and can cause an electric shock This is especially true for the very high voltage generated by the ozone generators It is dangerous to touch any parts inside the device In th...

Страница 4: ...arning icons They are located in the manual on the housing of the device or on other parts of the steriliser and are directly related to it Symbol Meaning Symbol Meaning Important safety instruction B...

Страница 5: ...pale cold skin sweat chills accelerated heart rate place them in an anti shock position on their back with their legs raised Despite a specially designed disinfection channel and limiting switch prot...

Страница 6: ...pain tearing itching redness avoid direct sunlight cool your eyes with a cloth dampened with cold water or glasses with cooling gel Do not scratch or rub your eyes If you have contact lenses it is ess...

Страница 7: ...s disease and increased ailments in people with coexisting respiratory and cardiovascular diseases In very high concentrations it is dangerous to live The effect of ozone on living organisms depending...

Страница 8: ...es avoid contact with eyes Wash your hands thoroughly with water and soap after providing medical assistance Inhalation If the airways come into contact with too high a concentration of ozone leave th...

Страница 9: ...eir indications are correct Improper measurements of the ozone sensor can lead to life threatening and health threatening effects as described in this chapter The ozone sensor is calibrated by the man...

Страница 10: ...dance with EN 136 and EN 14387 standards and in the skin protection equipment in accordance with the personal protective equipment table When the ozoning of the room is complete it must be thoroughly...

Страница 11: ...in the room including electronics it is also advisable to move them during disinfection The use of ozone sterilisation ozoning from time to time should not affect the destruction of objects in the roo...

Страница 12: ...ria and allergens using a coating applied to the filter medium In addition to particles the active carbon filter retains harmful gases and odours through adsorption In the UV C disinfection mode the U...

Страница 13: ...on dose 149 J m2 232 J m2 443 J m2 Ozone generator capacity 30 000 mg h 45 000 mg h 60 000 mg h Net weight 18 kg 20 kg 35 kg in sterilisation mode Name STERYLIS STERYLIS STERYLIS Model VS 400 1200 VS...

Страница 14: ...ts with other molecules breaking down their structures The effectiveness of ozone in eliminating odours lies in the fact that as a gas it is able to reach wherever there is air i e all the cracks nook...

Страница 15: ...pecially selected type and high power of the UV C light source with a selective emitted wavelength of 253 7 nm and the unique solution of the double quartz filter UV C lamp used in STERYLIS steriliser...

Страница 16: ...g manual 6 Safety data sheet ozone 7 UV C safety data sheet 8 Safety data sheet 9 Declarations of conformity in Polish and English 10 Two warning signs WARNING NO ENTRY STERILISATION IN PROGRESS as be...

Страница 17: ...front of the device The user panel consists of a main switch controls individual modes of operation failure and warning function keys three segment displays and a graphical display of the ozone diagr...

Страница 18: ...ONCENTRATION section segment display indicating the ozone concentration and the and SET buttons Alarm icons for failure of a system of ozone generators ozone sensor or UV C fluorescent lamps Main devi...

Страница 19: ...enter the standby mode and then display the corresponding error codes on the display in the O3 STERILIZATION section except for errors F201 and F202 which cause the PRE TEST to stop until the error is...

Страница 20: ...PRE TEST function and then it will start to light steadily and the device will start working in the selected mode of operation Confirmation of the disinfection mode is the operation of the front panel...

Страница 21: ...where the user can change the parameters of ozoning time and working ozone concentration 6 3 4 1 O3 sterilisation automatic mode To start O3 STERILIZATION in automatic mode press and hold the yellow S...

Страница 22: ...igh frequency inactive numbers light up constantly The PROG SET button changes the active digit to the next one on the right After entering the correct code confirm it by pressing the yellow START but...

Страница 23: ...During the ozone destruction process the LED backlight panel is activated When the safe concentration 0 1 ppm is reached the ozone safety light will light up After completion of the sterilization pro...

Страница 24: ...ower supply of the device should be disconnected as safely as possible preferably by means of an overcurrent switch in the switchgear or if necessary by unplugging the plug of the power cord from the...

Страница 25: ...generators 0 1 0 C150 Date setting day of the month 1 31 15 C151 Date setting month 1 12 7 C152 Date setting year 2020 2120 2020 C200 FIRMWARE revision read only FW_REV The individual parameters are c...

Страница 26: ...value of the electricity consumed is previewed using the d201 meter 6 8 Other functionality of the device Other remaining functionality of the device is described in subsections 6 8 1 6 8 2 and 6 8 3...

Страница 27: ...e sensor error has been detected during previous use of the unit the OZONE SENSOR TEST function can erase this error if the unit will function correctly during this function applies to the following e...

Страница 28: ...nd setting the current date The function is used to set the current time on the controller Selecting the SET TIME function displays the currently set time on the display in the OZONING TIME section To...

Страница 29: ...1 and 7 2 2 Do not let the filters clog and avoid their excessive operation After the service life of the UV lamps 9 000 operating hours in the steriliser they should be replaced This is related not...

Страница 30: ...t air filter cartridge towards the floor 4 Remove the air filter from the cartridge 5 Insert a new air filter into the cartridge making sure that the direction of the arrow on the cartridge and on the...

Страница 31: ...of the end point also results in a corresponding error on the display in the O3 STERILIZATION section If the end point violation stops the error is automatically reset The safety limiting switches pr...

Страница 32: ...s time has been exceeded will be displayed Note The replacement of UV C fluorescent tubes in STERYLIS sterilisers may only be carried out by the manufacturer s service or another authorised service ce...

Страница 33: ...STERILIZATION mode according to the instructions in section 6 3 4 9 Exit the room by closing the door behind you and make sure again that it is tightly sealed and that no one is in the room immediatel...

Страница 34: ...vice life of air filters F003 Exceeded service life of ozone generators F004 Exceeded service life of the ozone sensor F005 Exceeded expiry time of the ozone sensor F006 Air filters expiration date ex...

Страница 35: ...t blinking PRE TEST function is in progress for standard disinfection mode active errors Yellow pictograms O3 and flashing PRE TEST function for sterilization mode in progress no active errors Yellow...

Страница 36: ...use within certain voltage limits Directive 2014 30 EU on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Directive 2011 65 EU on the restriction of the u...

Страница 37: ...t with the service department Phone 48801080 257 E mail address serwis emiloo pl When contacting the service department you may be required to provide the data on the unit s nameplate Serial number Da...

Отзывы:

Похожие инструкции для STERYLIS VS-100/450