6.3 Starting the device and selecting the operating mode
The device can operate in four operating modes, the activation of which is forced by means of appropriate buttons on the
user panel. The following modes are available: Filtration, UV-C air disinfection – standard mode, UV-C air disinfection –
silent mode and O
3
sterilisation
To start the device, insert the power cord plug into an electrical outlet. The voltage required for proper operation is 230 V,
50 Hz. Then set the main switch of the device to ON.
When the device is started, it will be in standby mode, which means that it is ready for operation and waiting for a
command from the user. In the standby mode, the device indicates:
- The name of the AUTO or LOCH program (when the device is in locked mode) on the display in STERILIZATION O
3
section
and error and message codes (if any)
- On the display in the O
3
CONCENTRATION section: Value of the current ozone concentration when the concentration
exceeds 0.1 ppm / Characters - -.- - during the ozone sensor warm-up phase / In other cases, the display is inactive
Exiting the standby mode (for entering the access code, operating mode settings or device parameter settings) and
leaving the device idle for 60 seconds automatically returns to standby mode.
6.3.1 “Filtration” mode
To start the "Filtration" mode of operation, press and hold the START button located at the green filtering mode indicator.
The filter mode light will start blinking with low frequency (the device will check the proper functioning of the relevant
actuators - PRE-TEST function) and then it will start to light steadily and the device will start working in the selected
mode. If any errors are detected during the test, the device will display a tESt message, enter the standby mode and then
display the corresponding error codes on the display in the O
3
STERILIZATION section (except for errors F201 and F202,
which cause the PRE-TEST to stop until the error is corrected - the filter cartridge is inserted)
During operation in filter mode, the device indicates:
- Characters - - - - on the display in the O
3
STERILIZATION section or the LOCH message (when the device is in locked
mode) and error and message codes (if any)
- On the display in the O
3
CONCENTRATION section: Value of the current ozone concentration when the concentration
exceeds 0.1 ppm / Characters - -.- - during the ozone sensor warm-up phase / In other cases, the display is inactive
- Characters - - : - - on the display in the OZONING TIME section
The filtration mode can also be directly selected if the unit is in disinfection mode. To switch from disinfection mode to
filtration mode, press and hold the START button located at the green filtration mode light.
To interrupt the filtration process, press and hold the START button located at the green filtration mode light again.
6.3.2 “UV-C air disinfection - standard mode” operating mode
To start the "UV-C air disinfection - standard mode" operating mode, press and hold the START button located at the
green standard mode light. The standard mode light will start blinking with low frequency (the device will check the proper
functioning of the relevant actuators - PRE-TEST function) and then it will start to light steadily and the device will start
working in the selected mode of operation. Confirmation of the disinfection mode is the operation of the front panel
illuminated on the unit.
If any errors are detected during the test, the device will display a tESt message, enter the standby mode and then display
the corresponding error codes on the display in the O
3
STERILIZATION section (except for errors F201 and F202, which
cause the PRE-TEST to stop until the error is corrected - the filter cartridge is inserted)