Sterylis VS Manual
| 34
9.
Possible errors and messages
All possible errors and messages are indicated by 4-digit codes on the display in the O
3
STERILIZATION section. If more
than 1 error occurs simultaneously, the codes are displayed one after the other, the presentation time of each code is 2
seconds. The meaning of all errors and messages is described in the table below.
Code
Description
P003
Ozone process error - ozone concentration limit exceeded
P004
Ozone process error - threshold ozone concentration not reached
P005
Ozone process error - the ozone sensor was not properly heated
E101
UV-C sources malfunction - too low power consumption
E102
Fan operation error - too low power consumption
E103
Operating error of the ozone generators - too low power consumption
E201
Air filter error F1
E202
Air filter error F2
E300
Ozone generator error - relay / triac output shorted
E500
Ozone sensor error - reference voltage
E501
Ozone sensor error - incorrect reading
F001
Exceeded service life of the UV-C lamp
F002
Exceeded service life of air filters
F003
Exceeded service life of ozone generators
F004
Exceeded service life of the ozone sensor
F005
Exceeded expiry time of the ozone sensor
F006
Air filters expiration date exceeded
Important notes on the occurrence of codes from the above table:
Errors P003 and P004 are cleared when the device is restarted and when any mode of operation is activated.
Error P005 is cleared after the ozone sensor's warm-up time (default 30 minutes after starting the device)
Exxx errors are stored and cannot be deleted by restarting the device.
E500, E501, F004, F005 – If one of these errors occurs, the "Safe Ozone Concentration" light is always off.
Errors E201 and E202 should disappear when the filter cartridges are inserted back into the device again and
correctly.
Errors E101, E102 and E103 may be cleared by using the PRE-TEST and AUTO-TEST functions
Errors E500 and E501 may be cleared by using the OZONE-SENSOR-TEST function
Fxxx messages can be deleted by resetting the corresponding counters:
F001 – counter d102
F002 – counter d105
F003 – counters d108 and d111
F004 – counter d114
F005 – counter d117
F006 – counter d120
Error E300 is a critical error and can only be remedied after the fault has been fixed!