Sterylis VS Manual
| 22
instructions in 6.4) and a flashing signal generated by the warning beacon (regardless of the measured ozone
concentration). When the safe ozone concentration (0.1 ppm) is exceeded, the safe ozone concentration indicator will stop
illuminating.
The sterilisation process or the countdown to the start of sterilisation can be interrupted at any time by pressing any
button on the user panel. The device will then go into standby mode.
During the T
S
phase, the display in the OZONING TIME will show the characters - - . - -. During phases T
0
and T
D
, a
countdown of the sterilization time will take place, which is continuously updated on the basis of the current ozone
concentration.
During the ozoning process, the display in the O
3
CONCENTRATION section shows the current ozone concentration. If the
heating up of the ozone sensor is still in progress during the ozone treatment, the display shows the characters - -.- -.
After completion of the sterilization process, the device will switch off the ozone generators and after 30 minutes switch
on the UV-C fluorescent tubes to accelerate the ozone decomposition (ozone destruction). During the ozone destruction
process the LED backlight panel is activated. When the safe concentration (0.1 ppm) is reached, the ozone safety light will
light up.
After completion of the sterilization process the device will go into standby mode.
6.3.4.2 “O
3
sterilisation – manual” mode
To start O
3
STERILIZATION in automatic mode, press and hold the PROG.SET button in the O
3
STERILIZATION section on
the user panel. Enter the correct access code to enter manual mode.
Entering the access code:
The 4-digit code entered is shown on the display under O
3
STERILIZATION. The + and - buttons change the value by 1
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The
PROG.SET button changes the active digit to the next one on the right. After entering the correct code, confirm it by
pressing the yellow START button longer. If the code is incorrectly entered, the code XXXX or HHHH will be displayed and a
short beep will occur. In this case, the code must be re-entered.
After approving a correctly entered code, the sterilisation process parameters must be completed. The display in the O
3
STERILIZATION section will show one of the following programs: SEt.1, SEt.2 or SEt.3. These are three different options
for setting the parameters of the selected mode available to the user.
To select one of the above mentioned settings, press the + button or the - button in the O
3
STERILIZATION section to
scroll up or down the setting selection accordingly.
The settings of the currently selected program will be displayed in the O
3
STERILIZATION and OZONING TIME sections. To
change the settings of the preset ozone concentration, control the + and - buttons in the O
3
STERILIZATION section to
increase or decrease the value of the active digit by 1 respectively. The active digit is changed using the SET button. To
change the setting of the preset ozoning time, control the + and - buttons in the OZONING TIME section to increase or
decrease the value of the active digit by 1 respectively. The active digit is changed using the SET button.
To save the settings for a given program, press and hold the PROG.SET button in the O
3
STERILIZATION section. Save of
the settings will be confirmed by a short beep and a single blinking of the concentration and time displays.
To save the settings of all the programs and to proceed directly to the sterilization process (with SET 1 / SET2 / SET 3
selected), press and hold the yellow START button in the STERYLIZACJA O
3
STERILIZATION section. The display in the O
3
STERILIZATION section will show the selected type of sterilization, the display in the OZONING TIME section will start