Enhancement Param.
P806 (V)
Echo Strength
Use this feature to view the strength (in dB above 1
µ
V rms) of the echo selected to
base the distance calculation upon.
Values:
0 to 99
P807 (V)
Noise
Observe the average and peak ambient noise (in dB above 1
µ
V rms) being
processed.
The noise level is a combination of transient acoustic noise and electrical noise
(induced into the transducer cable or receiving circuitry itself).
Values:
x.y
x = average (-99 to 99), y = peak (-99 to 99)