P341
run time
The cumulative days that the application device has been in service. The
time is recorded once daily in a non-resetable counter. Periods of less
than 24 hr. are not recorded, nor accumulated.
P350
calibration security
This parameter provides additional security to the global lock (P000).
entry: 0 = view parameters, perform zero and span, no reset
of totalizer 1
1 = same as level 0, but cannot perform span
2 = same as level 0, but cannot perform zero and span
P359
factoring
Factoring is used as a method of calculating the value of the test
load (P017) to a new physical test weight.
entry: 0 = idle
1 = factor
Totalization is halted during the factoring procedure,
and resumed only upon return to the run mode.
P360
calibration duration
Sets the number of belt revolutions to use during a zero or span calibration.
enter number of belt revolutions, range 1 - 99
P367
direct zero
This parameter allows the zero reference count to be viewed
or entered directly.
Direct entry is intended for use when replacing software or hardware
and it is not convenient to perform an initial zero at that time.
PARAMETER DESC.
PL-531
65