Load Cell Balancing Parameters, P291 - P295
These parameters are used for verifying or balancing the load cells (2)
on the associated conveyor belt scale. Refer to Start Up\Balancing
for details and procedure for use of these parameters.
P291
calculator input 1
This register displays the count associated with the summation of
load cell A and B signals, when balancing the A and B load cells
of the associated belt scale.
P292
calculator input 2
This register displays the count associated with the load B signal,
when balancing the A and B load cells of the associated belt scale.
P293
calculator input 3
This register displays the count associated with the load cell B signal,
when balancing the A and B load cells of the associated belt scale.
P294
calculator input 4
This register displays the count associated with the summation
of load cell A and B signals, when balancing the A and B load
cells of the associated belt scale.
P295
load cell balance
Used in conjunction with balance calculator parameters (P291 - P 294),
this parameter displays the adjustment required to complete the load
cell balance procedure.
End of balancing parameters
PARAMETER DESC.
PL-531
64