Altering a program
You can alter all parameters which are identified as changeable in the
program charts. Other settings can only be altered by Miele Service.
A program setting is altered in two steps:
– First the wash blocks must be reallocated to the program or the
existing allocation confirmed again. Only allocated program blocks
can be parameterized.
– Then the program parameters can be altered.
Document all changes of factory settings in case of a subsequent
Service call.
Changing program parameters on a validated machine will
necessitate a renewed performance validation.
...
Program settings
Change program
Change program
Universal
Standard
Select the program you want to alter.
For more information see "Allocating wash blocks".