Program settings
102
Repeating if limit
value is exceeded
If the conductivity exceeds the limit value for the water intake or the
water drain, the measurement can be repeated.
If the limit value of the water intake is too high, the water is pumped
out and the water intake is repeated. if the conductivity at the end of
the rinse cycle is above the limit value for the drained water, the entire
final rinse is repeated.
If the value exceeds the limit again, the program will be cancelled with
a error message. The message will be displayed both on the display
and the cycle report.
...
Conductivity limit
Water intake
or
Water drainage
Number of repeats
Number of repeats
1
(0 - 1)
Adjustment is in increments of 1. The possible range is shown in the
bottom line.
Use the arrow buttons
(higher) and
(lower) to set the number of
repetitions.
Press
OK
to save the setting.
Calibrating the
conductivity
meter
The conductivity meter must be recalibrated by Miele Service at
regular intervals, e.g. during servicing. Sometimes calibration may
also be necessary outside the service cycle.
Conductivity module
requires calibration
If this message appears, contact the Miele Service Department.