Due to the high efficacy of the
Oxivario and Orthovario processes it
is important to lubricate joints and
contact parts on instruments pre
sterilisation. Please refer to the
instrument manufacturer for
guidance.
Warning and Safety
instructions
,
These warning and safety
instructions are in addition to those
given at the beginning of this
booklet.
~
H
2
O
2
solution must only be used in
the special containers and adapters
provided by the Ecolab or Dr. Weigert
companies.
~
Please observe safety instructions
(safety data sheets) provided by the
chemical agent manufacturer.
~
Be especially careful when handling
H
2
O
2
solution. It is an irritant chemical.
Observe all current safety instructions
pertaining to its use.
Wear protective gloves and goggles.
~
Empty containers must be disposed
of in accordance with the
manufacturer's instructions.
~
H
2
O
2
solution must not be mixed
with other chemicals. This could cause
a serious chemical reaction, e. g. the
release of harmful vapours or gases.
~
Only use tenside cleaning agents
supplied by Ecolab or Dr. Weigert.
Connecting the H
2
O
2
solution
container
The connection hose for the H
2
O
2
solution container is identified by a
black label. The connection hose is
supplied without a safety cap as the
container for hydrogen peroxide will
vary depending on supplier.
^
Connect the supplier's safety cap to
the connection hose (black).
^
Then connect this to the H
2
O
2
solution container.
^
Then start the
DOS2-FILL
programme.
Unlike with other chemical agents the
H
2
O
2
solution container must be fully
emptied before it is exchanged for a
new one.
Do not replace the H
2
O
2
solution
container with a new one until the
FILL
DOS 2 CONTAINER
message is
displayed. After connecting the new
container you must run the
DOS2-FILL
programme.
If the
CHECK DISPENSING SYSTEM 2
message appears you should check
the container as well as the dispensing
system. The programme will stop
automatically.
Special processes - OXIVARIO and ORTHOVARIO
74