Height
117.5 cm
Width
90 cm
Depth
Depth with door open
70 cm
126.5 cm
Weight (net)
180 kg
Voltage, connected load, fuse rating
See data plate
Mains cable
Approx. 1.8 m
Water temperature: Cold water
Hot and AD water
max. 20 °C
max. 70 °C
Static water pressure
Max. 1000 kPa (10 bar) pressure
Minimum flow pressure:
Cold/hot water inlet connections
AD water connection
100 kPa pressure
60 kPa pressure
Recommended flow pressure:
Cold/hot water inlet connections
AD water connection
0
250 kPa pressure
0
150 kPa pressure
AD-water connection (non-pressurised)
optional, 8.5 -50 kPa
Delivery head
min. 0.6 m, max. 1 m
Steam condenser
Approx. 2 - 4 l/min
Ambient temperature
5 °C to 40 °C
Relative humidity
Decreasing proportionately to
80 % for temperatures to 31 °C
50 % for temperatures to 40 °C
Altitude
Max. 1500 m
#
Degree of soiling
P2 (according to IEC/EN 61010-1)
Protection category (according to IEC
60529)
IP20 (Dust permeation)
Noise level in dB (A), Sound level LpA
during cleaning and drying phases
< 70
Test marks
VDE, radio interference suppression
!
mark
MDD-Guidelines 93/42/EWG, Class IIb
Manufacturer's address
Miele & Cie. KG, Carl-Miele-Str. 29,
33332 Gütersloh, Germany
#
If installed at altitudes above 1500 m the boiling point of the suds solution will be
lower. Disinfecting temperature parameters should be lowered and the holding
time increased (A
0
value). This must be done by a Miele authorised service
technician.
Technical data
71