– Large surface area filters are
enclosed in the kit supplied with the
machine for installing between the
stopcock and the inlet hose (see
illustration in Problem solving -
"Cleaning the filters in the water inlet").
The large surface area filter is made
from chrome nickel steel and can be
recognised by the matt top surface.
,
The inlet hoses must
not
be
shortened or damaged in any way
(see illustration).
See also the installation diagram
supplied.
AD water connection (pressurised)
ö
60 - 1000 kPa over pressure
The machine is supplied as standard
for connection to an AD water supply of
60 - 1000 kPa. If the water pressure
(flow rate) is below 150 kPa the water
intake time is automatically extended.
^
The AD pressure tested hose
(marked "H
2
O pur") with a 3/4"
threaded union must be connected to
the on-site AD stopcock for purified
water.
,
Note: If the machine is not going
to be connected to an AD water
supply, the cleaning and
disinfecting programmes will have to
be altered by a Miele Service
technician.
The inlet hose will remain on the rear
of the machine.
Optional AD water connection
(without pressure) 8.5 - 60 kPa
The appliance
must
be converted
(special accessory) (if this has not
already been done at the factory)
before connecting to a water supply
8.5 - 60 kPa. The feed pump must be
fitted by a Miele service technician
only.
When connecting an AD water
container (without pressure) the drain
connection must be at least the height
of the upper edge of the machine (see
"Installation instructions").
Plumbing
68