Switching on
^
Open the stopcocks (if turned off).
^
Press button
s
for at least 1.5
seconds.
The length of time for which the
s
button can be pressed is set at a
maximum of 10 seconds (see
Programming manual "System
functions/How long you can press the
On/Off button").
^
In operating levels A, B and D the
most recently selected programme
appears in the display and in
operating level C, instructions for
Automatic mobile unit recognition
is displayed.
The display will switch off
automatically after approximately 10
minutes when the washer-disinfector
is not in operation.
Press any button to switch the
display back on again.
Reset button
If the machine stops reacting to buttons
being pressed use the reset button.
The reset button is situated underneath
the electronic control unit. To operate it:
^
Open the DOS drawer.
^
Press the reset button.
This shuts down the profitronic unit and
then restarts it.
Starting the programme
See the Programme charts in the
Programming manual supplied with
the machine for detailed information
and important notes on the standard
Miele programmes.
,
When reprocessing medical
devices, any changes made to
programmes or dispensing systems
must be documented in a log book
kept with the machine (in
accordance with the Medical Device
Directive). The machine's cleaning
and disinfecting standards must
also be re-validated.
Operating levels A and B
A list of all the available programmes
will appear.
^
Select the programme required with
- e
and confirm with
OK
.
^
Press the Start button
h
.
The programme will proceed.
Operating level D
^
Select the
Programme overview
option and confirm your selection
with
OK
.
^
Select the programme required with
- e
and confirm with
OK
.
^
Press the Start button
h
.
The programme will proceed.
Operation