The type plate of the pressure sensor is located on the upper end of the cable and contains the following
specifications:
l
Item number
l
Measurement range (e.g. 0-10 m)
l
Type of pressure sensor (e.g. "9" for series PR-26D Ei)
l
Permissible media (e.g. "W" for fresh, waste and grey water)
Type plate of the pressure sensor
The correct specification of the type designation and serial number is important for all queries and spare part
orders. Only then can we process requests promptly and properly.
Note:
These operating instructions are part of the device and must be available to the user at all
times. The safety instructions contained therein must be observed.
WARNING:
It is strictly prohibited to disable the safety equipment or modify its mode of operation.
5.10 Installation of spare and wear parts
Be advised that spare and accessory parts that have not been supplied by the manufacturer have also not
been inspected or approved by the manufacturer. The installation and/or use of such products can possibly
have a negative impact on the specified constructional properties of the device. The manufacturer shall not be
liable for any damages that arise from the use of non-original parts and non-original accessory parts.
Note:
The use of spare and wear parts that are not approved by the manufacturer shall void the Ex
approval.
5.11 Storage of the product
To store the myDatalogGauge 3G , activate transport mode by removing the Activation module (BLE ATM)
from the sensor recess in the device. As part of this process, a connection to the myDatanet server will first
be established so that any data that has not yet been transferred to the server at this point in time is
transmitted. The data transmission and measurement are then deactivated. The antenna can be removed as
soon as the 2G/3G connection has been terminated - indicated by "Transport mode, no activation module
inserted" being shown on the display (see "Display" on page 76). You may need to activate the display again
by briefly holding the magnet (approx. 1 sec.) to the reed switch again (see "Reed switch" on page 75). The
BLE Gauge sensor modules do not need to be deactivated separately. Store all of the components of the
myDatalogGauge 3G (BLE & 3G Gateway ATEX and BLE Gauge sensor modules) in their original
packaging. Do not remove the protective armour during this process.
The myDatalogGauge 3G is placed in a very energy-saving mode by activating transport mode. However, it
may still occur that the batteries are fully discharged if the device is stored for very long periods. However, the
configuration and the most recently recorded data are always retained. Transport mode is deactivated again
32
Rev. 01
Содержание myDatalogGauge 3G
Страница 2: ......
Страница 10: ...2 2 BLE Gauge 10 Rev 01 ...
Страница 11: ...Chapter 2 Declaration of conformity Rev 01 11 ...
Страница 12: ......
Страница 13: ...Chapter 3 Ex certification Chapter 3 Ex certification 3 1 BLE 3G Gateway ATEX Rev 01 13 ...
Страница 14: ...14 Rev 01 ...
Страница 15: ...Chapter 3 Ex certification Rev 01 15 ...
Страница 16: ...16 Rev 01 ...
Страница 17: ...Chapter 3 Ex certification 3 2 BLE Gauge Rev 01 17 ...
Страница 18: ...18 Rev 01 ...
Страница 19: ...Chapter 3 Ex certification Rev 01 19 ...
Страница 20: ...20 Rev 01 ...
Страница 134: ......
Страница 140: ......
Страница 152: ......
Страница 153: ...Chapter 18 Document history Chapter 18 Document history Rev Date Changes 01 16 04 2021 First version Rev 01 153 ...
Страница 154: ......