Chapter 16 Troubleshooting and repair
16.2 Log entries and error codes
Log entry
Parameter
Description
Code
Plain text
Code
Plain text
1000 POWER ON
0
---
Restart following a power failure
4
---
Watchdog reset (e.g. because of an exception)
6
---
Reset was initiated by the device itself (e.g. in
event of firmware update)
##
--
Restart for another reason. There may be a
hardware problem if the "POWER ON" log entry
with a parameter code that is not equal to 0 or 6
is contained in the device log several times.
Contact the manufacturer in this case (see
"Contact information" on page 157).
1030 UV LOCKOUT
---
---
The device switches to energy saving mode and
terminates all of the operations as the
rechargeable battery or battery voltage is too
low. Only the charge controller, if present,
remains active.
1031 UV RECOVER
---
---
The rechargeable battery or battery voltage once
again suffices to guarantee reliable operation.
This is achieved by replacing the rechargeable
battery or battery pack . The device resumes
normal operation in accordance with the
configuration.
1034 CONTROLLER
UPDATE
##
---
Controller firmware update was completed
successfully
This entry is always duplicated in the device log.
In the first entry, the parameter specifies the
major version number (e.g. 3 for 03v011), while
in the second entry it specifies the minor version
number (e.g. 11 for 03v011).
1035 EXCEPTION
##
---
An internal system error was detected that
caused the device to restart. The parameter
specifies the type of system error. Contact the
manufacturer if the device log contains this error
with the same parameter code several times
(see "Contact information" on page 157).
1038 UV MODEM
LOCKOUT
---
---
The device deactivates the modem because the
rechargeable battery or battery voltage is too
low. A connection cannot be established now.
Rev. 01
143
Содержание myDatalogGauge 3G
Страница 2: ......
Страница 10: ...2 2 BLE Gauge 10 Rev 01 ...
Страница 11: ...Chapter 2 Declaration of conformity Rev 01 11 ...
Страница 12: ......
Страница 13: ...Chapter 3 Ex certification Chapter 3 Ex certification 3 1 BLE 3G Gateway ATEX Rev 01 13 ...
Страница 14: ...14 Rev 01 ...
Страница 15: ...Chapter 3 Ex certification Rev 01 15 ...
Страница 16: ...16 Rev 01 ...
Страница 17: ...Chapter 3 Ex certification 3 2 BLE Gauge Rev 01 17 ...
Страница 18: ...18 Rev 01 ...
Страница 19: ...Chapter 3 Ex certification Rev 01 19 ...
Страница 20: ...20 Rev 01 ...
Страница 134: ......
Страница 140: ......
Страница 152: ......
Страница 153: ...Chapter 18 Document history Chapter 18 Document history Rev Date Changes 01 16 04 2021 First version Rev 01 153 ...
Страница 154: ......