82
<Command>
:SWEep:CURRent<currentNR3/disc>
:SWEep:CURRent?
Function
:
It sets or queries the LEVEL to current mode, when the TYPE setup is FREQ.
Description
:
Set parameterThe value of current
AC RO 100Ω is {0.0002~0.02|MAXimum|MINimum}.
The value of current AC RO 25Ω is {0.0002~0.04|MAXimum|MINimum}.
Set syntax:SWEep:CURRent 1
:SWEep:CURRent 1M
:SWEep:CURRent 1MA
:SWEep:CURRent 1E-3
:SWEep:CURRent MAXimum
:SWEep:CURRent MINimum
Query syntax:SWEep:CURRent?
Return data 1.03(Format is in <NR3>)
When the LEVEL setup is not current mode, the data returns 9.9E37.
<Command>
:SWEep:TRIGger:MODE <REPeat|SINGle|0|1>
:SWEep:TRIGger:MODE?
Function
:
It sets or queries the TRIG.
Description
:
Set parameterREPeat,0 (repeat)
,
SINGle,1 (single)
Set syntax:SWEep:TRIGger:MODE SINGle
Query syntax:SWEep:TRIGger:MODE?
Return data REP | SING (Format is in <disc>)
<Command>
:SWEep:SPEEd {FAST|MEDium|SLOW|1|2|3}
:SWEep:SPEEd?
Function
:
It sets or queries the SPEED.
Description
:
Set parameterFAST/1, MEDium/2, SLOW/3
Set syntax:SWEep:SPEEd 1
:SWEep:SPEEd FAST
Query syntax:SWEep:SPEEd?
Return data FAST |MED | SLOW (Format is in <disc>)
<Command>
:SWEep:TRACe {A|B}
:SWEep:TRACe?
Function
:
It sets or queries which trace is on using.
Description
:
Set syntax:SWEep:TRACe A
Querysyntax:SWEep:TRACe?
Return data A | B (Format is in <disc>)