29
4.19.2 BIN METHOD
Equal:Set the upper and lower limit values for parameter, the Equal
mode averages each the comparison values automatically. The Limit
mode can select <ABS>, <
△
Abs> and <
△
%>.
Sequential:The Sequential mode specifies comparison limits as the
absolute measurement value. The Limit mode can select <ABS>, <
△
Abs> and <
△
%>.
Tolerance:The tolerance mode specifies comparison limits by the
deviation from the nominal value. The Limit mode can select<
△
Abs>
and <
△
%>.