69
<Command>
:MEASure:RANGe:DC {1|2|3|4|AUTO|HOLD}
:MEASure:RANGe:DC?
Function
:
It sets or queries the DC RANGE.
Description
:
Set parameter{1|2|3|4|AUTO|HOLD}
Set syntax:MEASure:RANGe:DC 2
:MEASure:RANGe:DC AUTO
Query syntax:MEASure:RANGe:DC
Return Data 2 (Format is in <NR1>)
<Command>
:MEASure:RANGe:AC {1|2|3|4|AUTO|HOLD}
Function
:
It sets or queries the AC RANGE.
Description
:
Set parameter{1|2|3|4|AUTO|HOLD}
Set syntax:MEASure:RANGe:AC 3
:MEASure:RANGe:AC AUTO
Query syntax:MEASure:RANGe:AC
Return data 3 (Format is in <NR1>)
<Command>
:MEASure:OIMPedance {100|25}
:MEASure:OIMPedance?
Function
:
It sets or queries the RO.
Description
:
Query syntax:MEASure:OIMPedance?
Return data 100|25 (Format is in <NR1>)
<Command>
:MEASure:COMParator:PARAmeter {1|2|3|4}
:MEASure:COMParator:PARAmeter?
Function
:
It sets or queries which parameters number is to be compared.
Description
:
Set syntax:MEASure:COMParator:PARAmeter 3 (After this command, every compared
setup is at this parameters number.)
Query syntax:MEASure:COMParator:PARAmeter?
Return data3 (Format is in <NR1>)
<Command>
:MEASure:COMParator:STATe {OFF|ON|0|1}
:MEASure:COMParator:STATe?
Function
:
It sets or queries the COMP open or not.
Description
:
:MEASure:COMParator:STATe?
Return dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command>
:MEASure:COMParator:MODE <ABSolute|DEViation|PERCent|0|1|2>
Function
:
It sets or queries the COMP MODE.
Description
:
Set parameterABSolute,0|DEViation,1|PERCent,2
Set syntax:MEASure:COMParator:MODE PERCent
Query syntax:MEASure:COMParator:MODE?
Return dataPERC (Format is in <disc>)