38
6.1.1 SWEEP DELAY
The delay time between each sweeping point can be set according to
the user’s need. When set at Sweep Delay Mode, it will affect the entire
measuring time.
NOTE:
The trigger delay time is set as 0-5000mS. If
exceeding the high limit value, it will be adjusted back
to 5000mS by the system automatically.
6.1.2 OUTPUT IMPEDANCE
The<100Ω>or<25Ω> can be selected according to the user’s demand.
The varied signal source output impedance will lead to the varied current
or the difference of measuring value. If selecting <25Ω>, then the AC
voltage range change to 10mV~1Vrms, and the current range is
200µA~40mArms.