HygroSmart HS3 User’s Manual
Michell Instruments
33
APPENDIX B
R/W Key
Value Description
R
Always readable
W
Always writeable
RC
Can only be read when REG_CUSTOMER_PASSWORD (register 35) contains 4660,
otherwise returns 0
WC
Can only be written when REG_CUSTOMER_PASSWORD (register 35) contains 4660,
otherwise write operation is ignored
Note 1 - Temperature Register
Value
Description
1
Temperature units = °C
2
Temperature units = °F
All parameters that sue temperature unit (temperature/dew-point/frost-point/wet bulb/output
settings/temperature correction table) will be automatically converted into the newly selected
temperature unit.
Note 2 - Customer Password
Writing
4460
to this register unlocks all registers marked as RC or WC.
Note 3 - Output 1 & 2 Type
Value
Quantity to output
1
Temperature
2
%RH
3
Dew Point
4
Frost Point
5
Water Activity
6
Absolute Humidity (g/m
3
)
7
Wet Bulb
8
Enthalpy
9
water vapour pressure over water (Int16)
10
water vapour pressure over ice (Int16)
See also notes 1 & 4.