HygroSmart HS3 User’s Manual
22
97516 Issue 5, July 2017
APPENDIX A
6 MAINTENANCE
6.1
Cleaning the Filter Element
It is necessary to keep the
fi
lter clean to ensure good air
fl
ow around the sensor element,
and to prevent a buildup of contamination which could cause harm to the sensor.
The cleaning procedure is as follows:
1.
Carefully remove the
fi
lter cap from the HS3 sensor.
Figure 28
Filter Cap Removal
2.
Flow distilled water through the
fi
lter element from the inside out.
Figure 29
Filter Element Cleaning
3.
Flow alcohol through the
fi
lter element from the inside out.
4.
Using compressed air, dry the
fi
lter element from the inside out.
If there is any doubt about the cleanliness of the
fi
lter after cleaning, a replacement can
be ordered - order code A000018.
Содержание Hygrosmart HS3
Страница 1: ...Hygrosmart HS3 Advanced RH Temperature Probe User s Manual 97516 Issue 5 July 2017 ...
Страница 32: ...HygroSmart HS3 User s Manual 24 97516 Issue 5 July 2017 APPENDIX A Appendix A Technical Specifications ...
Страница 48: ...HygroSmart HS3 User s Manual 40 97516 Issue 5 July 2017 NOTES ...
Страница 49: ......
Страница 50: ...http www michell com ...