HygroSmart HS3 User’s Manual
34
97516 Issue 5, July 2017
APPENDIX B
Note 4 - Output 1 & 2 Min/Max
The register type and divider for Output 1 Min and Output 1 Max are inherited from the parameters
selected in Output 1 Type. For example, if Water Activity is selected, then the type is UInt16, and the
divider is 100 as per the register map entry.
The register type and divider for Output 2 Min and Output 2 Max are inherited from the parameters
selected in Output 2 Type.
Register TYPE_A (Status Register)
Bit
Flag
0
Output 1 under range
1
Output 1 over range
2
Output 2 under range
3
Output 2 over range
4
Thermistor open-circuit
5
Thermistor short-circuit
6
RH element open-circuit
7
RH element short-circuit
8
Sensor communications error
9
Input volts too low
10
Logic volts too low
11
Temperature out-of-range
12
RH out-of-range
13
Not Used
14
Customer extended access
15
Diagnostic mode enabled
One or more bits are set to '1' to indicate an error or alternative mode of operation.
Register TYPE_B (Serial Number Registers)
Two consecutive 16-bit registers form a 32-bit unsigned integer, X, that stores the serial number. The
displayed serial number has the format HH-DDDDDD, where H is a hex digit & D is a decimal digit.
HH = (X & 0x0FF00000) >> 20
(displayed as 2 hexadecimal digits)
DDDDDD = X & 0x000FFFFF
(displayed as a decimal number)
Example: If REG_PROBE_SERIAL_HI = 0x0AA5 and REG_PROBE_SERIAL_LO = 0x464E, then the
probe serial number is displayed as "AA-345678"
Содержание Hygrosmart HS3
Страница 1: ...Hygrosmart HS3 Advanced RH Temperature Probe User s Manual 97516 Issue 5 July 2017 ...
Страница 32: ...HygroSmart HS3 User s Manual 24 97516 Issue 5 July 2017 APPENDIX A Appendix A Technical Specifications ...
Страница 48: ...HygroSmart HS3 User s Manual 40 97516 Issue 5 July 2017 NOTES ...
Страница 49: ......
Страница 50: ...http www michell com ...