HygroSmart HS3 User’s Manual
16
97516 Issue 5, July 2017
LOOP CHECKING SIMULATION
After Correction:
Figure 22
Probe readings post-adjustment
3
rd
Adjustment Point.
1.
Place the HS3-Probe together with a reference in a controlled and stable
environment of ± 70%RH.
2.
Check the readings of the HS3-Probe and the reference.
•
The Humidity reference is reading 69.6%RH.
•
The HS3-Probe reads 71.3%RH.
Figure 23
Probe readings prior to adjustment
3.
Find the nearest point in the RH Adjustment table (in this example 70) and
change it to the measured reference value (in this example 69.6%RH).