HygroSmart HS3 User’s Manual
30
97516 Issue 5, July 2017
APPENDIX B
B.3 Register
Map
Addr:
Gives the (base 10) address of each 16 bit register stored in the instrument.
R/W:
Refer to R/W key following this table
Type:
Indicates the data type of the register. For complex types, a letter is indicated which corresponds
to a description of the type found following this table.
Div:
Some register values with a
fi
xed number of decimal places are stored internally as integers,
which must then be divided by the number in this column to give the real value.
Addr Description
R/W
Type
Div Internal Name
Comment
0
Instrument
Modbus address
R/W
UInt16
REG_OWN_ADDR
1-32
Measured & Calculated Parameters
1
Temperature
R
Int16
10
REG_TEMPERATURE
-250.0C TO
250.0C
2
%RH
R
UInt16
10
REG_RH
0.0 to 100.0
3
Dew Point
R
Int16
10
REG_DEWPOINT
-364.0C to
364.0C
4
Frost Point
R
Int16
10
REG_FROSTPOINT
-364.0C to
364.0C
5
Water-Activity
R
UInt16
100 REG_WATER_ACTIVITY
0.000 to
1.000
6
Absolute
Humidity (g/m
3
)
R
UInt16
10
REG_ABS_HUMIDITY
0.0 to 6553.5
7
Wet Bulb
R
Int16
10
REG_WET_BULB
-364.0C to
364.0C
8
Enthalpy
R
Int16
10
REG_ENTHALPY
-3000.0 to
3000.0
9
water vapour
pressure over
water
R
UInt16
REG_WVP_WATER_INT
0 to 65535
10
water vapour
pressure over ice
R
UInt16
REG_WVP_ICE_INT
0 to 65535
11
water vapour
pressure over
water HI
R
IEEE754
single
REG_WVP_WATER_F_HI
12
water vapour
pressure over
water LO
REG_WVP_WATER_F_LO
13
water vapour
pressure over ice
HI
R
IEEE754
single
REG_WVP_ICE_HI
14
water vapour
pressure over ice
LO
REG_WVP_ICE_LO
15-31 Reserved
Miscellaneous
32
Status
R
TYPE_A
REG_STATUS
Содержание Hygrosmart HS3
Страница 1: ...Hygrosmart HS3 Advanced RH Temperature Probe User s Manual 97516 Issue 5 July 2017 ...
Страница 32: ...HygroSmart HS3 User s Manual 24 97516 Issue 5 July 2017 APPENDIX A Appendix A Technical Specifications ...
Страница 48: ...HygroSmart HS3 User s Manual 40 97516 Issue 5 July 2017 NOTES ...
Страница 49: ......
Страница 50: ...http www michell com ...