HygroSmart HS3 User’s Manual
Michell Instruments
17
MAINTENANCE
4.
Subtract the Humidity value of the HS3-Probe from the reference value
(69.6 – 71.3 = -1.7).
•
Add the outcome to the reference value (69.6 -1.7 = 67.9).
•
Enter this value in the corresponding
fi
eld of Corrected RH.
Figure 24
Editing the Adjustment Table
After Correction:
Figure 25
Probe readings post-adjustment