3.2. Remote control commands
785 DMP Titrino
90
&Info
Object
Description
Input range
Reference
& Root
:
Ã
Info
Current data
:
Ã
.Report
Transmission of formatted reports $G
3.2.2.92.
³ Ã
.Select
Report type
configuration, parameters,
³
smpl data, statistics, silo, calib
³
C-fmla, def, user method, full,
³
short, mplist, curve, derive, comb,
³
adj para,
scalc full, scalc srt, calc,
³
act dir, mem card, all, ff
ditto
³
Ã
.CalibrationData
pH calibration data
$G
3.2.2.93.
³ Ã
.Inp1
For measuring input 1
³ ³ Ã
.pHas
Asymmetry pH
0...7.00...
±
20.00
ditto
³ ³ Ã
.Slope
Slope of electrode
0...1.000...
±
9.999
ditto
³ ³ Ã
.Temp
Calibration temperature
-170.0...25.0...500.0
ditto
³ ³ Ã
.Date
Date of calibration
read only
ditto
³ ³ Ã
.ElectrodeId
Id of the calibrated electrode
read only
ditto
³ Ã
.Inp2
For measuring input 2, as for input 1
³ Ã
.Diff
For differential input, as for input 1
³
Ã
.Checksums
Checksums
$G
3.2.2.94.
³ Ã
.MPList
Checksum of meas.point list
read only
ditto
³ Ã
.ActualMethod
Checksum of current method
read only
ditto
³
Ã
.DetermData
Determination data
$G
3.2.2.95.
³ Ã
.Write
Read/write for several nods
ON, OFF
³ Ã
.ExV
Volume of Exchange/Dosing unit
read only/read+write
ditto
³ Ã
.MPList
Measuring point list
³ ³ Ã
.1
Measuring point 1
³ ³ ³ Ã
.Attribute
Attribute
read only/read+write
ditto
³ ³ ³ Ã
.X
X coordinate
read only/read+write
ditto
³ ³ ³ Ã
.Y
Y coordinate
read only/read+write
ditto
³ ³ ³ Ã
.Z1
Z1 coordinate
read only/read+write
ditto
³ ³ ³ Ã
.Z2
Z2 coordinate
read only/read+write
ditto
³ ³ ³
for each measuring point
³
Ã
.TitrResults
Titration results
³ Ã
.RS
Calculated results
³ ³ Ã
.1
1
st
result
³ ³ ³ Ã
.Value
Value
read only
3.2.2.96.
³ ³ Ã
up to 9 results
"Info", continuation
³ Ã
.EP
Endpoint
³ ³ Ã
.1
1
st
result
³ ³ ³ Ã
.V
Value
read only
ditto
³ ³ ³ Ã
.Meas
Measured value
read only
³ ³ ³ Ã
.Mark
Mark if more than 1 EP per window read only
³ ³ Ã
up to 9 EP's
³ Ã
.Var
Variables C4X
³ ³ Ã
.C40
Start measured value
read only/read+write
ditto
³ ³ Ã
.C41
Titration end volume
read only/read+write
³ ³ Ã
.C42
Titration time
read only/read+write