![Metrohm 785 DMP Titrino Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/metrohm/785-dmp-titrino/785-dmp-titrino_instructions-for-use-manual_1779564041.webp)
2.7. Result calculations
785 DMP Titrino
38
2.7 Result calculations
Formula entry, key <DEF>
Key <DEF> contains various inquiries for result
calculations and data output. The data of this key are
method specific and they are stored in the method
memory together with the method.
formula
:
Formulas for result calculations.
The display texts of the Titrino are shown to the left.
The values are the default values.
>formula
RS?
RS1=
RS1=EP1*C01/C00
Input of formulas
Enter formula number (1...9)
You can calculate up to 9 results per method.
Enter a number 1...9.
Input of formula
Example:
RS1=EP1*C01/C00
Enter formula by means of 3rd functions of keyboard.
Here you will find operands, mathematical operations
and parentheses. Operands require a number as an
identification. You can use the following operands:
EPX: EP's. X = 1...9
RSX: Results which have already been calculated with
previous formulas. X = 1...9.
CXX: Calculation constants. XX = 00...79.
Rules:
•
Calculation operations are performed in the
algebraic hierarchy: * and / and -.
•
Store formula with <ENTER>.
•
Calculation quantities and operands can be deleted
with <CLEAR> one by one.
•
To delete a complete formula press <CLEAR>
repeatedly until only RSX remains in the display.
Confirm with <ENTER>.
If a formula is stored with <ENTER>, result text,
number of decimals and result unit will be requested:
DEF
(
2
def
>formula
>silo calculations
>common variables
>report
>mean
>temporary variables