2.6. Parameters, key <PARAM>
785 DMP Titrino
24
2.6.2 Parameters for SET
SET1, SET2:
Control parameters for EP1 and EP2.
titration parameters
are valid for the global titration sequence.
stop conditions:
Parameters for the termination of the titration.
statistics
:
Calculation of mean values and standard deviation,
see page 41.
preselections
:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.
>SET1
EP at pH OFF
**titr.
dynamics OFF
**titr.
max.rate 10.0 ml/min
**titr.
Control parameters for EP1 or EP2, resp.
Preset EP1 (input range depends on the measured
quantity:
pH:
0...±20.00, OFF
U, Ipol:
0...±2000 mV, OFF
Upol:
0...±200.0 uA, OFF)
<CLEAR> sets "OFF".
If EP1 is "OFF", no further inquiries under SET1 appear.
Distance from EP where constant dosing should stop
and controlling begins. (control range, input range
depends on the measured quantity:
pH:
0.01...20.00, OFF
U, Ipol:
1...2000 mV, OFF
Upol:
0.1...200.0 uA, OFF)
<CLEAR> sets "OFF".
"OFF" means largest control range, i.e. low titration.
Outside the control range, dispensing is performed
continuously, see also page 30.
Maximum dosing rate (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
This parameter determines primarily the addition rate
outside the control range, see also page 30.
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
parameters
>SET1
>SET2
>titration parameters
>stop conditions
>statistics
>preselections