3.2 Remote control commands
785 DMP Titrino
81
*Parameter
Tree part "Parameters for SET"
Ã
.
SET1
Control parameters for EP1
³ Ã
.EP
Endpoint 1
depends on meas.quant. 3.2.2.28.
³ Ã
.UnitEp
Unit of endpoint
read only
ditto
³ Ã
.Dyn
Dynamics
depends on meas.quant. 3.2.2.29.
³ Ã
.UnitDyn
Unit of dynamics
read only
ditto
³ Ã
.MaxRate
Maximum dosing rate
0.01...10...150, max.
ditto
³ Ã
.MinRate
Minimum dosing rate
0.01...25.0...9999
ditto
³ Ã
.Stop
Titration stop
³ ³ Ã
.Type
Type of stop criterion
drift, time
3.2.2.30.
³ ³ Ã
.Drift
Stop drift
1...20...999
ditto
³ ³ Ã
.Time
Switch-off delay time
0...10...999, inf
ditto
³ ³ Ã
.StopT
Stop time
0...999 999, OFF
ditto
Ã
.
SET2
Control parameters for EP2, as for EP1
Ã
.
TitrPara
Titration parameters
³ Ã
.Direction
Titration direction
+, -, auto
3.2.2.31.
³ Ã
.XPause
Waiting time before start volume
0...999 999
3.2.2.32.
³ Ã
.Start V
Start volume
³ ³ Ã
.Type
Type of start volume
abs., rel., OFF
3.2.2.15.
³ ³ Ã
.V
Volume for absolute start volume
0...999.99
ditto
³ ³ Ã
.Factor
Factor for relative start volume
0...
±
999 999
ditto
³ ³ Ã
.Rate
Dispensing rate for start volume
0.01...150.0, max.
ditto
³ Ã
.Pause
Waiting time after start volume
0...999 999
3.2.2.16.
³ Ã
.ExtrT
Extraction time
0...999 999
3.2.2.33.
³ Ã
.MeasInput
Measuring input
1, 2, diff.
3.2.2.18.
³ Ã
.Ipol
Polarization current
0...1...
±
127
ditto
³ Ã
.Upol
Polarization voltage
0...400...
±
1270
ditto
³ Ã
.PolElectrTest
Test for polarized electrodes
ON, OFF
ditto
³ Ã
.Temp
Titration temperature
-170.0...25.0...500.0
3.2.2.19.
³ Ã
.TDelta
Time interv. for meas.acquisition
1...2...999 999
3.2.2.34.
³
Ã
.
StopCond
Stop conditions
³ Ã
.VStop
Stop volume
³ ³ Ã
.Type
Type of stop volume
abs., rel., OFF
3.2.2.20.
³ ³ Ã
.V
Volume for absolute stop volume
0...99.99...9999.99
ditto
³ ³ Ã
.Factor
Factor for relative stop volume
0...
±
999 999
ditto
³ Ã
.FillRate
Filling rate
0.01...150.0, max.
3.2.2.23.
³
Ã
.
Statistics
Statistics
³ Ã
.Status
Status of statistics calculation
ON, OFF
3.2.2.24.
³ Ã
.MeanN
No. of individual determinations
2...20
ditto
³ Ã
.ResTab
Result table
³ ³ Ã
.Select
original,delete n,delete all
ditto
³ ³ Ã
.DelN
Deletion of individual results
1...20
ditto
³
Ã
.
Presel
Preselections
³ Ã
.Cond
Conditioning
ON, OFF
3.2.2.35.
³ Ã
.DriftDisp
Display of drift during cond.
ON, OFF
ditto
³ Ã
.DCor
Drift correction
³ ³ Ã
.Type
Type of drift acquisition
auto, man., OFF
ditto
³ ³ Ã
.Value
Drift value for manual drift corr.
0.0...99.9
ditto
³ Ã
.IReq
Request of Id's after start
id1, id1&2, all, OFF
3.2.2.26.
³ Ã
.SReq
Request of smpl size after start
value, unit, all, OFF
ditto
³ Ã
.LimSmplSize
Limits for sample size
3.2.2.27
³ ³ Ã
.Status
Status of limit control
ON, OFF
ditto
³ ³ Ã
.LoLim
Lower limit
0.0...999 999
ditto
³ ³ Ã
.UpLim
Upper limit
0.0...999 999
ditto
³ Ã
.ActPulse
Output of a pulse
first, all, cond., OFF
3.2.2.28.