4.2.
Diagnosis
785 DMP Titrino
141
4 Cylinder code, date, time
•
Insert exchange unit or dummy to the dosing unit and put the burette tip into a collecting
receptacle.
•
Press
<0>
.
date/time
cylinder code
•
Press
<Enter>
.
date XX-XX-XX
XX:XX:XX
code:D0 XX ml
activated dosing unit
ml-code
•
Check date and time.
•
Check whether the ml-code does correspond with the exchange unit placed or not.
For the sake of completeness, different exchange units can be inserted to check their ml code.
If desired, the
exchange unit can be removed again. If no exchange unit is inserted display does not show the ml-code but
„check exchange unit !“.
•
Press
<Clear>
.
diagnose press key 0...9
5 Motor timer test
•
Press
<6>
.
motor-timer test
•
Press
<Enter>
.
pot.meter dV/dt
➜
➜
➜
➜
10?
•
Turn knob ‘
dVdt
’ to the right stop and press
<Enter>
.
Test
sequence:
a) In a first step, the frequency of the RC oscillator (analogue rate) is tested over a
period of 1 second.
b) In a second step, the frequency of the quartz oscillator (digital rate) is tested over a
period of 1 second.
c) If no error is found, after about 5 s it appears
motor-timer test o.k.
•
Press
<Clear>
.
diagnose press key 0...9
6 Analogue input test
•
Press
<7>
.