3.2. Remote control commands
785 DMP Titrino
100
3.2.2.16. Mode.Parameter.TitrPara.Pause
0...999 999
Parameters for DET, MET, SET: Pause time in s. Is waited off after the dis-
pensing of the start volume.
3.2.2.17. Mode.Parameter.TitrPara.MeasInput
1, 2, diff.
Mode.Parameter.TitrPara.Ipol
-127...1...+127
Mode.Parameter.TitrPara.Upol
-1270...400...+1270
Mode.Parameter.TitrPara.PolElectrTest
ON, OFF
Parameters for DET, MET, SET:
Selection of the measuring input; valid with measured quantities pH and U.
"diff." means differential amplifier, see page 158.
With Ipol, the inquiries for the polarization current in uA (Ipol) and
.PolElectrTest are valid.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry
in steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on changeo-
ver from the inactive state to an active state (titration or conditioning).
3.2.2.18. Mode.Parameter.TitrPara.Temp
-170.0...25.0...500.0
Parameters for DET, MET, SET: Titration temperature in °C. If a Pt100 or
Pt1000 is connected, the temperature is measured continuously and the pa-
rameter .Temp is updated.
The temperature is used for the temperature correction in pH measurements.
3.2.2.19. Mode.Parameter.StopCond.VStop.Type
abs., rel., OFF
Mode.Parameter.StopCond.VStop.V
0...99.99...9999.99
Mode.Parameter.StopCond.VStop.Factor
0...
±
999 999
Parameters for DET, MET, SET: Stop volume.
If an absolute stop volume (abs.) has been selected, the volume in mL is
valid.
A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size
*
factor
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.20. Mode.Parameter.StopCond.MeasStop
pH: 0...
±
20.00, OFF
U: 0...
±
2000, OFF
I:
0...200.0, OFF
Mode.Parameter.StopCond.UnitMStop
read only
Parameters for DET and MET: Stop when a measured value is reached. Entry
as pH value, in mV (with U and Ipol) and in uA (with Upol). The appropriate
unit can be viewed with .UnitMStop.
OFF means that the criterion is not monitored.