![Metrohm 785 DMP Titrino Скачать руководство пользователя страница 148](http://html1.mh-extra.com/html/metrohm/785-dmp-titrino/785-dmp-titrino_instructions-for-use-manual_1779564148.webp)
4.2.
Diagnosis
785 DMP Titrino
145
•
Press
<5>
.
RS232 test 1...2
8.1 RS232 test 1
•
Press
<1>
.
RS232 test-connector? 1
•
Insert the 3.496.8560 test plug in port ‘A1’.
•
Press
<Enter>
.
Test
sequence:
1. In case of an error an error message is displayed. If for example no test plug is
connected, error message
error 68
appears).
2. If no error is found, after about 5 s display shows
RS232 test 1 o.k.
•
Remove test plug.
•
Press
<Clear>
.
RS232 test 1...2
8.2 RS232 test 2
•
Press
<2>
.
RS232 test-connector? 2
•
Insert the 3.496.8560 test plug in port ‘A2’.
•
Press
<Enter>
.
Test
sequence:
1. In case of an error an error message is displayed. If for example no test plug is
connected, error message
error 68
appears)
2. If no error is found, after about 5 s display shows
RS232 test 2 o.k.
•
Remove test plug.
•
Press
<Clear>
.
RS232 test 1...2
•
Press
<Clear>
.
diagnose press key 0...9