Single tests in PAT testing mode
73
Result screens:
Examples of Insulation LN-P measurement results
Displayed results:
Main result ............. Insulation resistance LN-P
Note
:
Consider displayed warnings before starting measurement!
6.3.6 Welding circuit leakage current
The purpose of this test is to determine the sum of all leakage currents flowing from the
welding outputs W1 or W2 to earth.
Welding circuit leakage current menu
Test parameters for welding circuit
current measurement
LIMIT
Maximum current
[3.50 mA, 5.00 mA, 10.00 mA, none]
TIME
Measuring time
[5 s, 10 s, 30 s, 60 s, 120 s, none]
Key (F3)
CHG-OFF Change between L-N on a single phase test socket of 3-phase adapter is
disabled.
CHG-ON
Change between L-N on a single phase test socket of 3-phase adapter is
enabled.
Test circuit and measurement procedure for welding circuit leakage current
measurement
For more information refer to chapter
Measurements according to IEC/ EN
60974-4
, paragraph
Welding circuit leakage current
in 3-phase adapter
Instruction manual.