MI 3321 MultiservicerXA
Autotest sequences
98
Visual test menu
Keys:
PASS (F1)
Confirms that the visual inspection passed.
SKIP (F4)
Skips visual test.
FAIL (F5)
Ends the autotest sequence.
7.6.2
Continuity of the protective circuit
Measurement is described in chapter
6.3.1 Continuity of the protective circuit
.
If the Continuity test fails or was skipped other tests will not be carried out because of
safety.
Keys:
START
Starts the Continuity measurement.
Proceeds with the next continuity measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
HELP (F2)
Displays the Continuity test help screens.
REPEAT (F3)
Repeats the Continuity measurement.
SKIP (F4)
Skips Continuity resistance measurement.
END (F5)
Ends the autotest sequence.
Notes
:
Compensation of test leads resistance should be performed before doing autotests.
For compensation of test leads Description in chapter 6.2.1.1
Compensation of test
leads resistance
can be used as reference.
7.6.3
Insulation resistance (supply circuit to protective circuit)
Measurement is described in chapter
6.3.2 Insulation resistance (supply circuit to
protective circuit)
. If this insulation test fails or was skipped other tests will not be carried
out because of safety.
Keys:
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single