MI 3321 MultiservicerXA
Instrument description
15
It is recommended to additionally check the DUT before
proceeding with the test!
Leakage LN-PE high.
Are you sure to
proceed? (Y/N)
Dangerous leakage current (higher than 3.5 mA) will flow if
power would be connected to DUT.
Select
YES
or
NO
with Y or N key.
Proceed with testing only if all safety measures have been
taken. It is recommended to perform a thorough earth bond
test on the PE of the DUT before proceeding with the test.
Leakage LN-PE too
high.
Dangerous leakage current (higher than 20 mA) will flow if
power would be connected to the DUT.
Determine and eliminate the problem before proceeding!
Leakage LN-PE or
S/C1-PE too high!
Are you sure to
proceed? (Y/N)
Dangerous leakage current (higher than 20 mA) would flow if
power were connected to the DUT.
Select
YES
or
NO
with Y or N key.
Proceed with testing only if all safety measures have been
taken.
It is recommended to perform a thorough earth bond test on
the PE of the DUT before proceeding with the test.
Measurement aborted!
Differential leakage too
high!
An exceptionally high leakage current (higher than about
5 mA) was measured through PE test terminal or PE socket
during the pre-test or test.
Leakage test S/C1-PE
too high!
An exceptionally high leakage current (higher than about
5 mA) was measured through S/C1 test terminal during the
pre-test.
External voltage on
test socket too high!
DANGER!
-
Voltage on mains test socket or LN/PE terminals
is higher than approximately 20 V (AC or DC)!
Disconnect the DUT from the instrument immediately and
determine why external voltage was detected!
External voltage on
S/C1 too high!
DANGER!
-
Voltage on test probe (S/C1) is higher than
approximately 25 V (AC or DC)!
Disconnect the test probe from the DUT and determine why
external voltage was detected!
Test was skipped for
safety!
Instrument skipped the required test because of a failed
previous test.