MI 3321 MultiservicerXA
Single tests in PAT testing mode
62
Clamp current menu
Test parameters for clamp current measurement
LIMIT
Maximum current
[0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25
mA, 2.50 mA, 3.00 mA, 3.50 mA, 5.00 mA, 9.00 mA]
TIME
Measuring time
[5 s, 10 s, 30 s, 60 s, 120 s, none]
Test circuit for clamp current measurement
Connecting current clamp to the instrument
Clamp current measurement procedure
Select the
Clamp current
function.
Set test
parameters.
Connect
the current clamp to the instrument (see figure above).
Embrace
wire(s) that has to be measured with current clamp.
Press the
START
key for measurement.
Examples of clamp current measurement result
Displayed results:
Main result ............. clamp current