12
Setting the parameters
12.1
Preparation
It is possible to change various settings directly on the meter. The 4 buttons located
underneath the front screw-on cover on the meter index head can be used for this
purpose, see section 11.2, page 21.
To reach the buttons, proceed as follows:
•
Remove the seal between the front and rear screws on the cover for the electronic
meter index.
•
Loosen the screws on the front of the meter index head and unscrew the cover.
•
Remove the 3 screws for the display cover.
•
Change the relevant parameter using the 4 buttons, see Fig. 10.
•
After changing the parameter, screw on the display cover and then the cover for the
electronic meter index head.
•
Fix the cover in place using the connection screws.
12.2
Changing the primary address
Set the desired primary address using the 4 buttons.
Action
Display
Input/output
1.
Press INC, then
SET.
PASS_0000
User password 1
•
No entry is required.
2.
Press SET.
002
_n
Primary address (0–255)
•
Change the digit with INC
•
Change the position with SHT
3.
Press RST.
SPA_0000
User password 2
•
Enter 1111 using INC (digit) and
SHT (position)
4.
Press RST.
EEP_SuCC
Data has been saved successfully
EEP_Err
Error when saving
29