Changing the turning range
The meter has a rotatable display with a turning range of 180° (A). The angle of rotation
can be changed to the other 180° (B) by turning the base by 180°.
1.
Loosen the 3 radial stud screws (1) on
the base of the meter index head.
2.
Slightly lift up the meter index head
until the positioning pin (3) has been
completely pulled out of the drilled
hole (2).
3.
Turn the meter index head by 180° in a
clockwise direction and lower it again
so that the pin is now seated in the
drilled hole on the opposite side (2).
4.
Tighten the stud screws (1) again by
hand.
5.
Make a note of the turning direction
so that if the meter index head is
adjusted again you can avoid turning
it in the same direction and possibly
damaging the cable connections.
1
Stud screws
2
Positioning pin
3
Drilled hole
14