EN
10
14. Replacement of the collection container
Turn off the main switch of the unit!
Prepare a new collection container and take out the enclosed
disinfectant bag!
Wear protective gloves!
Remove the lid of the treatment unit.
17
Turn the locking bracket of the amalgam separator upwards,
holding the separator firmly.
18
Remove the amalgam separator by pulling it forwards out of
the unit and place it on a level, non-slip surface.
19
Hold the collection container firmly, and twist the top counter-
clockwise until the marking under the locking bracket and the
yellow marking on the collection container are aligned. Lift
up the top.
20
If the pump filter is soiled, remove it, clean over a drip tray,
and place back onto the pump housing.
21
Fit the top onto a new collection container in such a manner
that the yellow markings on the collection container and un-
der the locking bracket are aligned. Twist the top clockwise as
far as it will go. Check whether all the supporting clips of the
container are properly locked into place.
22
Grease the seal on the supporting element with vaseline.
23
Carefully insert the amalgam separator into its support, and
close the locking bracket.
Switch on the main switch of the treatment unit.
If the collection container has been replaced properly, the amalgam se-
parator runs for a short period of time and the external display indicates
„ready for operation“ (signal 1, green illuminated).
If the collection container has not been replaced properly, then signal 2
starts flashing on the external display, and a buzzing sound is audible. In
this case, turn off the main switch and carefully repeat the procedure as
described above (refer to pt. 5 & 7).
Replacement of the collection container
Practice personnel, technicians