![MessKonzept FTC320 Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/messkonzept/ftc320/ftc320_operating-manual_1775514007.webp)
Page 6 of 43
1.2
Warning Notices
• The operator must ensure that the above classification and marking on the device is suitable for
the application.
• The cables must be connected voltage-free. Do not disconnect while energized!
• Permissible ambient temperature (when used as intended): -5 to +50 °C.
• A fixed wall mounting must be provided. The device is only to be mounted in areas with low
mechanical hazards, such as analysis rooms, in other words, rooms in which no heavy work is
performed.In areas with high mechanical hazards, such as workshops, transport routes, produc-
tion facilities for heavy equipment or construction sites, etc., additional protection against external
damage is necessary, e.g., by means of a protective cabinet.
• The device may only be used in an area with pollution degree 1 (according to DIN 60664-1).
• Avoid accumulation of dust on the device! Regular inspection of the housing and cables with
regard to mechanical damage is mandatory. In atmospheres that strongly attack plastics, the
front foil must also be checked for cracks and discoloration, especially in the area of the buttons.
If damage occurs, the device must be taken out of operation and the manufacturer must be
contacted.
• Warranty expires if the housing is opened.
• The unit and the cables must be effectively protected against damage and against UV light (pro-
tective roof for outdoor installation).
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf