Page 33 of 43
7.1.1
Set Offset Gas Concentration
Offset Gas
set:
0.00000
new:
0.00000
[%]
ESC/
OK
Calibrate
Offset Gas
Gain Gas
H
2
2.58 %
Calibration
Alarm Setup
H
2
2.58 %
Figure 7.2: offset gas setup menu
Before performing a calibration, the concentrations of the used test gases have to be set. In the sub-
menu “Offset Gas“ the used concentration of the offset gas has to be entered. Select the submenu
“Offset Gas“ by pressing the <RIGHT> key and press <ENTER> to open the menu for the numerical
entry. Enter the offset calibration gas concentration by changing each digit to the desired value. The
position of the cursor in the number is moved by the <RIGHT> key, the value of the digit (0-9 or . for
the decimal separator) at the current position of the curser is changed with the <UP> key. When the
correct value is set, move the cursor to “OK“ and confirm with <ENTER>.
7.1.2
Set Gain Gas Concentration
Gain Gas
set: 100.0000
new: 100.0000
[%] ESC
/
OK
Calibrate
Offset Gas
Gain Gas
H
2
2.58 %
Calibration
Alarm Setup
H
2
2.58 %
Figure 7.3: gain gas setup menu
Prior to a two point calibration, the gain gas concentration has to be set in the submenu “Gain Gas“.
This menu is operated analogously to the “Offset Gas“ menu described in Section 7.1.1.
7.1.3
Offset Calibration
Offset Cal.
Apply Gas
0.00
[%]
ESC/
OK
Stability
[%] ESC
/
OK
Actu. 0.00
Targ. 0.00
Sampling 10s
for Offset
Calibration
Please Wait
Save Calib.
Targ.
0.00
Actu.
0.00
[%]
ESC/
OK
Gain
Calibration?
------------
YES/
NO
Apply
Process Gas
Actu.
0.00
ESC/
OK
Calibrate
Offset Gas
Gain Gas
H
2
2.58 %
Calibration
Alarm Setup
H
2
2.58 %
Figure 7.4: menu path if only offset is calibrated
After opening the menu “Calibrate“, you are asked to apply the offset test gas. The third line of the
display shows the test gas concentration as set in the “Offset Gas“ menu (see Section 7.1.1). Please
calibrate using the same/a similar flow rate as in your process measurement situation. Confirming with
“OK“ leads to the menu “Stability“. The second line contains the set test gas concentration (“Targ.“),
the third line the measured concentration (“Actu. “) using the current (unchanged) calibration. Before
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf