Requirements for Electrical Connectors
Before using the device make sure that the power supply is in accordance with the
specifications of the device and that all electric connections correspond to the information
given in this manual.
The FTC320 is a device of protection class III. For power supply a source with PELV specification (Pro-
tective Extra Low Voltage) according to EN 60204-1 must be used. See also Section 3.4.2 "Ground".
The potential-free relay contacts must also be monitored with a power supply unit with PELV specifica-
tion.
3.4.2
Ground
ERROR
MAINT REQ
POWER
power supply
PELV complying
with EN 60204-1
+24V GND
A
B
C
GAS OUT
GAS IN
Figure 3.3: Grounding the FTC320
To comply with EN 60204-1 and to ensure your device’s function, the device has to be installed such
that the power supply (PELV) is connected to protective earth (PE) with its ground conductor, see
Figure 3.3. The shielding of cables A, B and C should be connected to functional ground. Dependent
on the circumstances, gas inlet and gas outlet can be grounded in addition. Connections to ground
should be made with short low-resistant cables of large diameter.
3.4.3
Data exchange via serial interface (RS-232)
The serial interface, often called UART (Universal Asynchronous Receiver Transmitter), is based on the
RS-232 standard. The point-to-point data transmission is carried out via the two TxD- (Transmit Data)
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf