continuing, wait for a sufficient running-in time to evacuate possible disturbing gases from the device.
Only continue if the value of the actually measured concentration reaches a final value (you might ob-
serve some signal noise around a constant value). To start the calibration sampling, select “OK“ and
press <ENTER>. The sampling phase of 10s is started. Based on the average measured concentra-
tion, the new offset is determined such that the currently measured (and calibrated) value accords with
the given test gas concentration within the specifications of the device. Repeat the calibration in case
the measured concentration after calibration is not in agreement with the test gas concentration. By
selecting “ESC“ and pressing <ENTER>, the offset calibration is repeated. With “OK“ the calibration is
confirmed as correct and the new offset value is saved.
The following menu offers the option to proceed with the gain calibration. Choosing “YES“ leads to
the gain calibration menu described below. It is highly recommended to select “NO“ at this point. The
gain value is very stable over time. Faulty gain calibration may worsen the device’s performance. After
quitting the calibration, you are asked to apply process gas again, giving you time to restore your typical
process measurement environment while error warnings are deactivated and relays stay frozen.
7.1.4
Gain Calibration
Save Calib.
Targ. 100 .00
Actu. 100 .00
[%] ESC
/
OK
Gain Calib.
Apply Gas
100.00
[%]
ESC/
OK
Stability
Targ. 100 .00
Actu. 99
.98
[%] ESC
/
OK
Sampling 10s
for Gain
Calibration
Please Wait
Gain
Calibration?
------------
YES
/NO
Apply
Process Gas
Actu. 100 .00
ESC/
OK
Figure 7.5: menu path of the gain calibration
The menu structure is designed such that the gain calibration procedure is only accessible after an off-
set calibration. The steps of the gain calibration correspond to the offset calibration described above (of
course now using a different gas concentration, as set up in the ”Gain Gas” menu, see Section 7.1.2).
Again, please remember to wait for a sufficient running-in time to evacuate possible disturbance gases.
The gain calibration can be canceled in any sub-menu by selecting “ESC“. If gain calibration is can-
celed, the previous slope of your calibration line remains unchanged.
7.2
Alarm Setup
Individual process alarms can be activated and configured for the selected measurand/gas in the alarm
setup accessible from the measurand related menu. Limit values and hysteresis can be set. Only one
alarm can be set per gas.
A triggered alarm is signaled by a flashing of the measured variable on the display. Alarms can be used
to switch relays.