If a reset to Factory Settings is performed without saving the calibration parameters
beforehand, a new calibration might be necessary.
If the device, against the given warnings, was reset to “Default Settings“ the parameters
relevant for the proper operation are overwritten by default values. With these values the
device will not work properly. The correct parameters (contact Messkonzept) have to be
written to the device again, see Section 6.5.1 or the manual of SetApp which you can find
on www.messkonzept.de.
6.5.4
Test of Relays, Analog Outputs and Connections
Parameter
Access Modes
Reset Funct.
I/O Test
Relays
Current Loop
Analog Out
Rel 1 open
Rel 2 open
Rel 3 open
DIN low
Relays
Current Loop
Analog Out
I-Out Const
set: 10.0000
new: 10.0000
[mA]
ESC
/OK
Relays
Current Loop
Analog Out
Analog Out 1
Analog Out 2
AO 1 Voltage
set: 0.00000
new: 0.00000
[V]
ESC
/OK
Analog Out 1
Analog Out 2
AO 2 Voltage
set: 0.00000
new: 0.00000
[V]
ESC
/OK
Figure 6.11: I/O Test menu
The “I/O Test“ menu provides the opportunity to set the following properties to a defined status in order
to test subsequently connected equipment:
• Relay 1 (Rel 1) (open/closed)
• Relay 2 (Rel 2) (open/closed)
• Common Relay (Rel 3) (open/closed)
• Current of current loop