Page 24 of 43
6.2.1
Parameter Menu
0.30
[%]
H2/N2
Parameter
Errors
Compound Raw
set: 3000.00
------------
P001 OK
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Figure 6.2: parameter menu for advanced diagnosis
The configuration of the FTC320 is defined by a list of internal parameters. The “Parameter“ menu gives
read-only access to these parameters. This may help an experienced user to diagnose malfunctions
caused by wrong settings. The parameter menu allows you to scroll through the entire parameter list.
Contact Messkonzept for detailed information on the listed parameters. The first display line contains
the name of the parameter, in the second line the parameter value is shown. The last line shows
the parameter index. To move forward in the list, press <ENTER>, to move backwards, press <Up>.
To leave the parameter menu mark <OK> by pressing <RIGHT> and confirm with <ENTER>. Some
parameters can be changed in the expert setup, see Section 6.5 for more information.
6.2.2
Errors
0.30
[%]
H2/N2
Parameter
Errors
No Errors
___
------------
Reset All?
NO/YES
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Figure 6.3: error menu
When the device is in the measuring mode or calibrating, several parameters are continuously checked
for plausibility. The parameters checked, the list of possible errors and the ranges defining a plausible
value can be found in Chapter 8 "Appendix: System Errors".
Errors are indicated by a flashing red light on the front panel. The menu “Errors“ gives access to the
list of current errors . “No Errors“ indicates that no error is pending, for pending errors the name of the
error is given. Pressing <ENTER> proceeds to the next pending error and <UP> to the previous error.
To leave the error menu press <RIGHT> once or twice to select ”NO” or ”YES” and the <ENTER> key
to confirm your selection. An error reset will affect maintenance-warnings only - all other errors and
warnings are continuously monitored and automatically reset when the error no longer persists.
Defective devices must be disconnected from the process!
This applies for apparent
damages of the device such as physical damages but also in the case of unclarified
malfunctions in the operation. Separate the device from the process pneumatically (both
gas inlet and gas outlet) and remove the power supply from the device.
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V02.pdf